Authors: A. Portu,I. Postuma,M.A. Gadan,G. Martin,M.S. Olivera,S. Altieri,N. Protti,S. Bortolussi,
Journal: Applied Radiation and Isotopes
https://doi.org/10.1016/j.apradiso.2015.07.022
Abstract: An inter-comparison of three boron determination techniques was carried out between laboratories from INFN-University of Pavia (Italy) and CNEA (Argentina): alpha spectrometry (alpha-spect), neutron capture radiography (NCR) and quantitative autoradiography (QTA). Samples of different nature were analysed: liquid standards, liver homogenates and tissue samples from different treatment protocols. The techniques showed a good agreement in a concentration range of interest in BNCT (1-100. ppm), thus demonstrating their applicability as precise methods to quantify boron and determine its distribution in tissues.